HD-DEN-16487_Dialin Config Helper Commands Are Not Supported With The Latest HD And DG Builds
HD-DEN-16487_Dialin Config Helper Commands Are Not Supported With The Latest HD And DG Builds
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when tr...

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when treatment is resumed.

HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

Done.

Done.

Align declaration comments as a block on the right.

Align declaration comments as a block on the right.

HD-DEN-15792_HD Software Design Description
HD-DEN-15792_HD Software Design Description
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

dryPressureTestsCompleted is already in this header. dryPressureTestsCompleted was added to setupForSelfTestStop() header.

dryPressureTestsCompleted is already in this header. dryPressureTestsCompleted was added to setupForSelfTestStop() header.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add dryPressureTestsCompleted to header.

Add dryPressureTestsCompleted to header.

Updated to handle any active alarm if we havent passed pressure tests. I also modified the logic to where this valve change should only happen in d.s.p.t.

Updated to handle any active alarm if we havent passed pressure tests. I also modified the logic to where this valve change should only happen in d.s.p.t.

Fixed. Thanks!

Fixed. Thanks!

What if we got some other alarm during pressure test states? Would it be better to check dryPressureTestsCompleted?

What if we got some other alarm during pressure test states? Would it be better to check dryPressureTestsCompleted?

Put constant on other side of comparison operator. We only put constant first when using "==".

Put constant on other side of comparison operator. We only put constant first when using "==".

HD-DEN-16164_FW Pre Treatment Pressure Test Failures Vent TO Saline Bag
HD-DEN-16164_FW Pre Treatment Pressure Test Failures Vent TO Saline Bag
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.