Merge branch 'DEN-9906-hd_dev-flow-measurement-changes' into develop

  1. … 3 more files in changeset.
DEN-9906: Fixed target rinseback volume published after rinseback stopped prematurely. Allow alarm response resume if in pause state of Tx End sub-mode.

  1. … 1 more file in changeset.
DEN-9906: Added rinse flag to drain commands

  1. … 2 more files in changeset.
DEN-9906: Fixed rinseback t/o countdown ref.

DEN-9906: DG Defs merge.

  1. … 1 more file in changeset.
Merge branch 'DEN-9906-hd_dev-flow-measurement-changes' into develop

  1. … 1 more file in changeset.
DEN-9906: Send new general event msg for HD start, all op mode changes. Fixed rinseback countdown timer issue DEN-10154.

  1. … 2 more files in changeset.
DEN-9906: Drain reservoir first during post-treatment

Merge branch 'DEN-9906-hd_dev-flow-measurement-changes' into develop

  1. … 5 more files in changeset.
Merge branch 'DEN-7802-dg_dev-dialysate-generation' into develop

  1. … 1 more file in changeset.
Merge branch 'DEN-7802-dg_dev-dialysate-generation' into staging

DEN-9906: Prevent transition from rinseback to recirc if rinseback not completed. Max additional rinseback volume set to 300 mL.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed dialysate temp check for treatment end sub-mode.

Removed dialysate temp check for treatment end sub-mode.

DEN-7802: Removed dialysate temp check for treatment end sub-mode

I think we do want to check dialysate temperature during treatment, but maybe not in every sub-mode. For example, I think in treatment end sub-mode we have the dialyzer bypassed (no longer performi...

I think we do want to check dialysate temperature during treatment, but maybe not in every sub-mode. For example, I think in treatment end sub-mode we have the dialyzer bypassed (no longer performing dialysis) so there is no need to worry about the dialysate temp anymore. That is the only exception I can think of at the moment.

DEN-9906: Delayed power on AC check to allow voltage monitor time to see 24V. Fix FPGA voltage ranges.

  1. … 4 more files in changeset.
DEN-9906: Added tentative blood flow estimation logic for testing. Added HD power-off function. Power off HD if no AC power on startup. Added t/o to rinseback stop state. Fixed valve commands in post-treatment mode.

  1. … 12 more files in changeset.
DEN-7802: Renamed re-circ mode to gene mode

  1. … 1 more file in changeset.
Merge branch 'DEN-9906-hd_dev-flow-measurement-changes' into develop

  1. … 1 more file in changeset.
DEN-9906: Added filtering for inline pressure sensors.

  1. … 2 more files in changeset.
DEN-9906: Extended rotor reset t/o from 10 to 20 sec. Added turn off of trimmer heater in blood prime. Fixed merge issue w/ integrity POST test. Split treatment recirc t/o alarm into an alarm and a warning.

  1. … 1 more file in changeset.
HD-DEN-7802_DG-DEV Dialysate Generation
HD-DEN-7802_DG-DEV Dialysate Generation
DEN-9906: Added a couple trimmer heater off commands to ensure heater does not stay on when DPi pump is not running.

  1. … 2 more files in changeset.
DEN-7802: Removed test success status on integrity test

  1. … 1 more file in changeset.
DEN-8886: second local merge from staging. Fixed merge conflicts.

  1. … 9 more files in changeset.