If going back to start state, set pump track closed required, but not door closed. If skipping dry self-tests, set both to be required closed. Also, this s/w config looks like it's setup wrong. Fo...

If going back to start state, set pump track closed required, but not door closed. If skipping dry self-tests, set both to be required closed.

Also, this s/w config looks like it's setup wrong. For release build, we want to go back to start state.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

transition and resume part of stopped state should set pump track required so we don't have to keep setting it here every 50 ms.

transition and resume part of stopped state should set pump track required so we don't have to keep setting it here every 50 ms.

Should set door and pump track required here before we move on to next state.

Should set door and pump track required here before we move on to next state.

Should set pump track closed required, but not door here.

Should set pump track closed required, but not door here.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Do we need an else where we reset 10 second start time (if latch opened)?

Do we need an else where we reset 10 second start time (if latch opened)?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Created handleDrySelfTestStartState() that counts settle time once the latch is determined closed.

Fixed. Created handleDrySelfTestStartState() that counts settle time once the latch is determined closed.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Only checking if BP is running when are in treatment mode.

Fixed. Only checking if BP is running when are in treatment mode.

Updated to 600.

Updated to 600.

Should this be moved into "if" body? Doesn't appear to need setting until we resume.

Should this be moved into "if" body? Doesn't appear to need setting until we resume.

Should this 10 seconds only start once latch is locked?

Should this 10 seconds only start once latch is locked?

Not sure exception for treatment stop state still makes sense. We can be running BP in that state now depending on alarm. Should still alarm on occlusion.

Not sure exception for treatment stop state still makes sense. We can be running BP in that state now depending on alarm. Should still alarm on occlusion.

So R2 fill was 1100 and you say it ended up at 500. Would be 800 if keep same at 1100. If we change to 600, would end around 300 as you say - I think that's right.

So R2 fill was 1100 and you say it ended up at 500. Would be 800 if keep same at 1100. If we change to 600, would end around 300 as you say - I think that's right.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

HD-DEN-14011_Occulsion Alarm When Reseating
HD-DEN-14011_Occulsion Alarm When Reseating
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.