DEN-14561: fixes to pre-tx fill cycles.

What does flush complete have to do with draining here?

What does flush complete have to do with draining here?

We don't want to signal here if we don't have an air pump.

We don't want to signal here if we don't have an air pump.

Done.

Done.

Seems like this flush tracking logic should be done in fill complete state (after fill, not before).

Seems like this flush tracking logic should be done in fill complete state (after fill, not before).

Add blank line between declaration and code.

Add blank line between declaration and code.

Another way to handle making sure we get past s.t. dry state is to not set startNormalFill on res until op mode state machine gets to s.t. dry state. Then don't need to change this if to consider p...

Another way to handle making sure we get past s.t. dry state is to not set startNormalFill on res until op mode state machine gets to s.t. dry state. Then don't need to change this if to consider pre-tx state.

DEN-14561

Reservoir selection update.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14549: Added alarm threshold to occlusion alarm data for logging. Only check for sufficient Heparin when in pre-tx mode. Addressed code review comments.

  1. … 2 more files in changeset.
Fixed.

Fixed.

Done, initial drain of the reservoirs, prior to syncing with the self-test.

Done, initial drain of the reservoirs, prior to syncing with the self-test.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Add alarmActionResumeReceived.

Add alarmActionResumeReceived.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14561

Partial fix for fill volumes pretreatment

We are asking DG to set the currently active reservoir as the active reservoir? So no change?

We are asking DG to set the currently active reservoir as the active reservoir? So no change?

Add spaces around (). And literal should go to right unless using == operator.

Add spaces around (). And literal should go to right unless using == operator.

Alignment of comment.

Alignment of comment.

I don't see this being used anywhere. Remove.

I don't see this being used anywhere. Remove.

HD-DEN-14718_Dry Self Test Failure: Air pump failing to reach target venous pressure within 30 seconds.
HD-DEN-14718_Dry Self Test Failure: Air pump failing to reach target venous pressure within 30 seconds.
Merge branch 'DEN-14718-dry-self-test-failure' into develop

DEN-14718: Increased air pump to 60s timeout. Fixed air pump not turning off when alarm 231 is triggered in self test state.

DEN-14549: Addressed bugs DEN-13769 and DEN-14231.

DEN-14561 Corrected air pump state.

After a dry self test failure, turned off the air pump.