DEN-13154: Ensure alarm system has control of alarm lamp in fault mode in case fault occurred before lamp test completed. Do not request DG active alarms on startup if DG is not communicating.

  1. … 1 more file in changeset.
SyringePump.c was included in this code review in error. The repositories associate the correct files with each branch.

SyringePump.c was included in this code review in error. The repositories associate the correct files with each branch.

Extra blank line.

Extra blank line.

Extra blank line.

Extra blank line.

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Skip synchronization with dialysate circuit priming.

DEN-13154: Fix handling of cartridge installed dry s.t. per DEN-14770.

HD-DEN-13154_Heparin Syringe Empty Position Incorrect
HD-DEN-13154_Heparin Syringe Empty Position Incorrect
HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off
HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off
DEN-14763 added code to send the results of the chem flush sample to DG

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
HD-DEN-14790_Alarm 199 Occurs Intermittently During Blood Prime And Treatment
HD-DEN-14790_Alarm 199 Occurs Intermittently During Blood Prime And Treatment
DEN-14790: Changed dialysate recirculation rate from 100 to 250 mL/min to add margin for minimum flow for heating.

  1. … 1 more file in changeset.
HD-DEN-14689_Messaging Bugs_HDFW
HD-DEN-14689_Messaging Bugs_HDFW
DEN-14273: Prevent certain DG alarms if AC power is lost (DEN-14550). Fix problem recovering from no cartridge installed alarm (DEN-14770).

  1. … 3 more files in changeset.
CORRECTED IN CODE WALKTHROUGH

CORRECTED IN CODE WALKTHROUGH

DEN-14273: Addressed code review comment.

Done.

Done.

Merge branch 'DEN-14589-DIAL-211' into develop

  1. … 5 more files in changeset.
#define for 0

#define for 0

HD-DEN-14273_Gui Invalid Treatment Parameters Are Not Highlighted Red
HD-DEN-14273_Gui Invalid Treatment Parameters Are Not Highlighted Red
DEN-14273: Added new check to treatment params validation - fail check if a Heparin pre-stop setting is given w/o Heparin bolus/dispense.

Fixed merge conflicts.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14561-2: fix to last.