Merge branch 'staging' into DEN-14689-messaging-bugs

  1. … 1 more file in changeset.
Fixed?

Fixed?

To answer this and comment below: above_high_safety_range = temp > 46 above_low_safety_range = temp > 42 above_target_temp = target + 4 below_target_temp = target - 4 below_low_safety_range = temp...

To answer this and comment below:

above_high_safety_range = temp > 46
above_low_safety_range = temp > 42
above_target_temp = target + 4
below_target_temp = target - 4
below_low_safety_range = temp < 33

I did not see any specification regarding to separate or not to separate. Original functionality had separate triggers. Respective PRS's state OR between the two triggers.
I think it would be clear to for us in terms of debugging to see which alarm is triggered since on the low side, temp - 4 would be lower than 33, (31).

Wording may be unclear. Im up for suggestions to help clarify the alarm differences.

Fixed. Thanks!

Fixed. Thanks!

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

This should be U32.

This should be U32.

This should be U32.

This should be U32.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Should we check for stall in bolus & continuous states?

Should we check for stall in bolus & continuous states?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

HD-DEN-14906_Dialyzer DE Prime
HD-DEN-14906_Dialyzer DE Prime
Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14834 updated the code to control the primary heater

  1. … 2 more files in changeset.
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

  1. … 1 more file in changeset.
DEN-13154: Further fix to DEN-14814 re: do not require syringe removal to re-do preload operation.

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14834 changed one of the pre-treatment target fills from 600 to 750mL Changed the reservoir's upper limit in treatment from 1950 to 1900mL

  1. … 1 more file in changeset.
Fixed.

Fixed.

Fixed.

Fixed.

Done.

Done.

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Address comment

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. D...

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm.
And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm.
Did Systems want to separate these? My understanding was not to separate.

Add blank line before return.

Add blank line before return.