Fixed merge conflicts from local merge from staging.

  1. … 4 more files in changeset.
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

FALSE will be determined by our automagic script to be ACK_REQUIRED in the excel sheet. This will cause discrepancies between the readers of that sheet and what the FW does.

FALSE will be determined by our automagic script to be ACK_REQUIRED in the excel sheet. This will cause discrepancies between the readers of that sheet and what the FW does.

Fixed.

Fixed.

Variable is also updated by receipts of DG flow broadcasts. I believe it is overridable on DG side.

Variable is also updated by receipts of DG flow broadcasts. I believe it is overridable on DG side.

Concern was originally just for retract due to highest speed. We ended up reducing retract speed to reduce likelihood of stalls. Still a good idea to check for it though and since retract, preload ...

Concern was originally just for retract due to highest speed. We ended up reducing retract speed to reduce likelihood of stalls. Still a good idea to check for it though and since retract, preload and seek all share same relatively high speed, these are the ops that have the check. For bolus and continuous, should not stall at these speeds and would catch with rate alarm if it did.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Merge branch 'staging' into DEN-14689-messaging-bugs

  1. … 1 more file in changeset.
Fixed?

Fixed?

To answer this and comment below: above_high_safety_range = temp > 46 above_low_safety_range = temp > 42 above_target_temp = target + 4 below_target_temp = target - 4 below_low_safety_range = temp...

To answer this and comment below:

above_high_safety_range = temp > 46
above_low_safety_range = temp > 42
above_target_temp = target + 4
below_target_temp = target - 4
below_low_safety_range = temp < 33

I did not see any specification regarding to separate or not to separate. Original functionality had separate triggers. Respective PRS's state OR between the two triggers.
I think it would be clear to for us in terms of debugging to see which alarm is triggered since on the low side, temp - 4 would be lower than 33, (31).

Wording may be unclear. Im up for suggestions to help clarify the alarm differences.

Fixed. Thanks!

Fixed. Thanks!

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

This should be U32.

This should be U32.

This should be U32.

This should be U32.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Should we check for stall in bolus & continuous states?

Should we check for stall in bolus & continuous states?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

HD-DEN-14906_Dialyzer DE Prime
HD-DEN-14906_Dialyzer DE Prime
Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces