DEN-14860 Fix config bug. If Dry tests disabled, got stuck in Prime because DG Allow Flush/Fill signals not set.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Updated signalAllowDGFlushFills(),

signalAllowDGFillRes[12]() to be sure that the signal to continue occurs even when both thr air pump and dry tests are disabled.

Add timeWaitToFillMS to header.

Add timeWaitToFillMS to header.

DEN-14906: Added delay before resuming dialyzer re-prime to allow DPi to stop to prevent pressure spike.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
DEN-14888: Fix issue with initializing start/final reservoir volumes.

DEN-14906: Handle UF pause and auto-resume for dialyzer re-prime.

I think we need to re-visit this. Could impact dialysate heating negatively.

I think we need to re-visit this. Could impact dialysate heating negatively.

Merge branch 'staging' into DEN-14763-dg_hd_dev-hd_dg_dvt-update-part-10

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14763 addressed the code review comments

  1. … 1 more file in changeset.
Merge branch 'DEN-14906-dialyzer-de-prime' into develop

Done

Done

DEN-14906: Restored 30 minute interval for dialyzer re-prime. Reduced re-prime time from 40 to 35 seconds. Fixed so treatment time is paused during dialyzer re-prime.