DEN-14906: Fixed isues that arose when disabling re-prime feature.

  1. … 1 more file in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed.

Fixed.

Merge DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off

Cherry picked merge, due to staging/develop conflicts

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand...

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand new alarm (triggered 0 ms ago).

What's the significance of changing to U32? Is this being used as a timer?

What's the significance of changing to U32? Is this being used as a timer?

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Integrated current changes in other branches to address apparent conflicts.

Should venPres < venLowLimit be done outside if statement?

Should venPres < venLowLimit be done outside if statement?

DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 5 more files in changeset.
Auto stash before merge of "DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off" and "staging"

Merge branch 'staging' into DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off

DEN-14906: restored 30 min interval for dialyzer re-prime.

DEN-14906: Increased pump speed during dialyzer re-prime and reduce time in re-prime state in attempt to address over-current alarm.

HD-DEN-14999_Alarm Table Updates
HD-DEN-14999_Alarm Table Updates
DEN-14860 Fix config bug. If Dry tests disabled, got stuck in Prime because DG Allow Flush/Fill signals not set.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Updated signalAllowDGFlushFills(),

signalAllowDGFillRes[12]() to be sure that the signal to continue occurs even when both thr air pump and dry tests are disabled.

Add timeWaitToFillMS to header.

Add timeWaitToFillMS to header.

DEN-14906: Added delay before resuming dialyzer re-prime to allow DPi to stop to prevent pressure spike.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2