Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Warning (13 min) removed since no longer alarming at 15 min.

Warning (13 min) removed since no longer alarming at 15 min.

Fixed.

Fixed.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Updated all headers.

Updated all headers.

Add timeOfLastUICheckin and uiDidCommunicate.

Add timeOfLastUICheckin and uiDidCommunicate.

Does the new code result in one less TRUE comparison? Can rinsebackTimerCtr ever be greater than RINSEBACK_DONE_WARNING_TIME?

Does the new code result in one less TRUE comparison? Can rinsebackTimerCtr ever be greater than RINSEBACK_DONE_WARNING_TIME?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed.

Fixed.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand...

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand new alarm (triggered 0 ms ago).

What's the significance of changing to U32? Is this being used as a timer?

What's the significance of changing to U32? Is this being used as a timer?

Should venPres < venLowLimit be done outside if statement?

Should venPres < venLowLimit be done outside if statement?

HD-DEN-14999_Alarm Table Updates
HD-DEN-14999_Alarm Table Updates
Add timeWaitToFillMS to header.

Add timeWaitToFillMS to header.

HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
I think we need to re-visit this. Could impact dialysate heating negatively.

I think we need to re-visit this. Could impact dialysate heating negatively.