Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14999 updated the code based on the alarms table

  1. … 1 more file in changeset.
DEN-14999-2: Removed alarm escalation feature.

  1. … 1 more file in changeset.
Isn't the ALARM_ID_HD_AC_POWE_LOST already cleared?

Isn't the ALARM_ID_HD_AC_POWE_LOST already cleared?

Interested in return value?

Interested in return value?

Comment alignment.

Comment alignment.

Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 1 more file in changeset.
DEN-14830: Reworked dialysate temperature logic so persistence triggers do not cancel each other out. Fixed air pump not turning off when dry self tests are user stopped.

  1. … 1 more file in changeset.
DEN-14999-2: prevent treatment resumption from rinseback/recirc states if an alarm that blocks resume has been triggered previously in the treatment.

  1. … 2 more files in changeset.
DEN-14999-2: Removed alarms 214, 215 and 217.

HD-DEN-14999_Alarm Table Updates Part 2
HD-DEN-14999_Alarm Table Updates Part 2
Merged staging

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-14860 Merge

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Merge DEN-14860 to staging

  1. … 2 more files in changeset.
DEN-14999: Removed redundant trigger for DG restarted alarm.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

I don't know. I believe check in msg should have no payload, but it usually does have a u32 of unknown purpose. I have found that the very first check-in after power up has no payload according to ...

I don't know. I believe check in msg should have no payload, but it usually does have a u32 of unknown purpose. I have found that the very first check-in after power up has no payload according to candump and this was causing alarm 46. Until UI sorts this out, I am not going to check payload.

What is the other flavor of UI check in message. If it is a different size, shouldn't it be its own separate message?

What is the other flavor of UI check in message. If it is a different size, shouldn't it be its own separate message?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14999: Removed obsolete/redundant alarms. Addressed code review comment.

  1. … 1 more file in changeset.