Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15353: explicit float casting of treatment duration parameter when calculating max volume.

DEN-15451 Fix re-enable of DPi on alarm resume.

Merge branch 'staging' into DEN-15196-DIAL-258

compiler fixes

implemented 4th level state changes to op status event. fixed some state transitions to display the correct current state.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Merged staging to DEN-15359.

HD-DEN-15196_DEN-15353_and_DIAL-258
HD-DEN-15196_DEN-15353_and_DIAL-258
DEN-15359: Misc fixes to art/ven pressure limit changes.

  1. … 2 more files in changeset.
Merge branch 'DEN-15359-usability-arterial-and-venous-pressure-alarm-limits' into develop

  1. … 1 more file in changeset.
DEN-15359: Fix s/w fault bug. Change so that dialysate flow change will also kick pressure limit s.m. back to stabilization state.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-15359: Addressed DEN-15445 by reducing flush fill volume from 500 to 400 mL to prevent leak out of DG port when cartridge is being connected.

DEN-15404 added the RO permeate sample command

  1. … 1 more file in changeset.
Done.

Done.

Merge branch 'DEN-15359-usability-arterial-and-venous-pressure-alarm-limits' into develop

  1. … 1 more file in changeset.
Thanks, that works too. RESOLVED IN CODE WALKTHROUGH

Thanks, that works too.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Add stabilizationStartTimeMs to header.

Add stabilizationStartTimeMs to header.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Can't keep in message x47 as f/w will not have calculated new min/max yet. I made it so message x9 will broadcast immediately after user changes pressure limit windows.

Can't keep in message x47 as f/w will not have calculated new min/max yet.
I made it so message x9 will broadcast immediately after user changes pressure limit windows.