DEN-15371 changed the pre-treatment target fill flow

HD-DEN-15371_Usability Fluid Leaks From Dialysate Port Before Cartridge Installation Sequence
HD-DEN-15371_Usability Fluid Leaks From Dialysate Port Before Cartridge Installation Sequence
Fixed merge conflicts from staging.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'staging' into DEN-15196-DIAL-258

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DIAL-258: added clear of 4th level state once we finish wet self tests.

fixed. set to no sub state (0xFF) once we transition to pretx recirc

fixed. set to no sub state (0xFF) once we transition to pretx recirc

Looks like this is only being set in wet self-tests substate of pre-tx prime. Once we leave wet self-tests, should this 4th level state be set back to zero or something?

Looks like this is only being set in wet self-tests substate of pre-tx prime. Once we leave wet self-tests, should this 4th level state be set back to zero or something?

Merge branch 'DEN-15317-usability-hd-door-open-alarm-during-post-treatment' into develop

  1. … 2 more files in changeset.
DEN-15317: Moved post-tx homing to ensure it happens in all cases.

DEN-15317: Addressed saline bolus issue as part of DEN-15359.

HD-DEN-15317_Usability HD Door Open Alarm During Post Treatment
HD-DEN-15317_Usability HD Door Open Alarm During Post Treatment
Merge branch 'DEN-15317-usability-hd-door-open-alarm-during-post-treatment' into develop

DEN-15317: Addressed DENF-15317.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-15451_Dpi Pump Does Not Resume After Door Opened Alarm Cleared IN Pre TX
HD-DEN-15451_Dpi Pump Does Not Resume After Door Opened Alarm Cleared IN Pre TX
FIXED IN CODE WALKTHROUGH

FIXED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Merge branch 'DEN-15451-dpi-pump-does-not-resume-after-door-opened-alarm-cleared-in-pre-tx' into develop

DIAL-258: code review changes