DEN-15404, DEN-15252, DEN-15485 work in progress. Implementing code to handle cleaning modes in HD standby mode

HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
DEN-15404, DEN-15252 added code the HD standby to transition to the corresponding cleaning mode

Merge branch 'staging' into DEN-15367-hd_dg_test_configuration_setup

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15314: Removed door closed requirement in post tx for patient disconnect confirm.

DEN-15314 changed the direction of the test configuration

Merge branch 'DEN-15314-usability-update-alarm-text-to-not-use-acronyms' into develop

  1. … 1 more file in changeset.
DEN-15314: Added stall check to syringe prime operation. Extended dialysate bypass prime timeout and added a little histeresis to exit criteria for bypass prime state.

  1. … 1 more file in changeset.
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
DEN-15314 added a test configuration to be able to do a one minute treatment

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15371 changed the pre-treatment target fill flow

HD-DEN-15371_Usability Fluid Leaks From Dialysate Port Before Cartridge Installation Sequence
HD-DEN-15371_Usability Fluid Leaks From Dialysate Port Before Cartridge Installation Sequence
Fixed merge conflicts from staging.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'staging' into DEN-15196-DIAL-258

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DIAL-258: added clear of 4th level state once we finish wet self tests.

fixed. set to no sub state (0xFF) once we transition to pretx recirc

fixed. set to no sub state (0xFF) once we transition to pretx recirc