Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-15278-blood-stop-air-trap-alarm-declares-during-saline-bolus-treatment' into develop

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Add bloodLeakPeristenceCtr.

Add bloodLeakPeristenceCtr.

DEN-15278: no countdown timer if alarm 165 (blood sitting too long) is active.

DEN-15367 addressed the code review comments

  1. … 3 more files in changeset.
Done

Done

Done

Done

Done

Done

That is true I fixed it.

That is true I fixed it.

Updated the function header.

Updated the function header.

Done

Done

Done

Done

Done

Done

Done

Done

Done

Done

Done

Done

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 3 more files in changeset.
Add bubbleClearEnded.

Add bubbleClearEnded.

DEN-15440 added code to partial drain the reservoirs in patient connect state machine in pretreatment

DEN-15367, DEN-15404 added a message to receive the message from DG that DG is ready to dispense more fluid in RO permeate sample

  1. … 3 more files in changeset.
Make 2nd condition similar to 1st ( != TRUE ) so any value that is not 1 will take normal path. And should this be an OR (not AND)?

Make 2nd condition similar to 1st ( != TRUE ) so any value that is not 1 will take normal path. And should this be an OR (not AND)?

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declarations and code.

Blank line between declarations and code.

Fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-15278: Skip pressurize/depressurize bubble clear portion of blood side cartridge prime if wet cartridge test config is set.