DEN-15556: Fix to last.

DEN-15556: Stabilize pressures after UF rate change.

  1. … 3 more files in changeset.
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
DEN-15612: Removed unnecessary change.

Fixed. Thanks!

Fixed. Thanks!

DEN-15621: code review comments.

DEN-15556: Addressed issues for DEN-15556 re: not repeating used cartridge test and openeing VBA after door closed after cartridge install to prevent saline from leaking into the line.

The 2 verify states are single pass. I don't think it makes sense to return to them. They either pass or you go to fault mode and so you wouldn't even be here - so I think you can assume you passed...

The 2 verify states are single pass. I don't think it makes sense to return to them. They either pass or you go to fault mode and so you wouldn't even be here - so I think you can assume you passed.
For first displacement verify state, I think you should go to second displacement setup state.
For second displacement verify state, I think you should go to complete state.

HD-DEN-15716_Skipped Treatment Start State While Transitioning From Pre Treatment TO Treatment
HD-DEN-15716_Skipped Treatment Start State While Transitioning From Pre Treatment TO Treatment
DEN-15621: implemented wet self recovery. Priming no longer goes to Priming pause state when in wet self tests.

HD-DEN-15556_FW Usability Pre TX Used Cartridge Failure
HD-DEN-15556_FW Usability Pre TX Used Cartridge Failure
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15763: Addressed code review comments.

  1. … 1 more file in changeset.
currentReservoirVolume is local. Others done.

currentReservoirVolume is local. Others done.

Done.

Done.

Add currentReservoirVolume, minimumReservoirVolume, steadyVolumeCount

Add currentReservoirVolume, minimumReservoirVolume, steadyVolumeCount

Add flowFilterTimerCtr, ufCalcTimerCtr

Add flowFilterTimerCtr, ufCalcTimerCtr

Merge branch 'DEN-15763-dg-restarted-alarm-107-cannot-be-cleared' into develop

DEN-15763: increased timeout period for dialysate prime state of pre-tx.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Update header: dopRateCorrectionOffset, offsetPWMDutyCyclePct, dopRateCorrectionOffset. dopControlSignal, dialOutPumpControlMode, dialOutPumpState

Update header: dopRateCorrectionOffset, offsetPWMDutyCyclePct, dopRateCorrectionOffset. dopControlSignal, dialOutPumpControlMode, dialOutPumpState

Merge branch 'DEN-15763-dg-restarted-alarm-107-cannot-be-cleared' into develop

  1. … 1 more file in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15763: Addressed DEN-15779.