Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'staging' into DEN-15716-skipped-treatment-start-state-while-transitioning-from-pre-treatment-to-treatment

Merge branch 'DEN-15788-pbo-ramps-up-to-400-mmhg-during-dialysate-circuit-priming' into develop

DEN-15788: change DPo control during dialysate/dialyzer prime state to begin immediately instead of after fixed prime time.

HD-DEN-15628_Blood Leak Sensor Zero During Wet Self Test Improvement
HD-DEN-15628_Blood Leak Sensor Zero During Wet Self Test Improvement
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused

force pumps to stop in recirc/treatment mode, on a manual stop.

HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused
HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused
DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Retrigger blocked, alarms directly blocked during stop

  1. … 2 more files in changeset.
Merged staging

  1. … 2 more files in changeset.
Merge branch 'DEN-15788-pbo-ramps-up-to-400-mmhg-during-dialysate-circuit-priming' into develop

  1. … 1 more file in changeset.
DEN-15788: Added 10 sec venous pressure filter and publish venous pressure readings with 10 sec filter. Use 10 sec venous pressure filter for setting pressure windows after stabilization. Fix proposed UF rate and Tx duration after a UF volume change. Add simple pressure control to DPo speed during dialysate/dialyzer prime state to prevent pressure build up.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-15628 changed the blood leak zero sequence counter from Z to G since Z is repeated twice in each zero sequence

  1. … 1 more file in changeset.
This part of code has been removed.

This part of code has been removed.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merged DEN-15660

  1. … 1 more file in changeset.
HD-DEN-15788_Pbo Ramps UP TO 400 Mmhg During Dialysate Circuit Priming
HD-DEN-15788_Pbo Ramps UP TO 400 Mmhg During Dialysate Circuit Priming
Merge branch 'staging' into DEN-15660-test_configuration_treatment_recovery

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15556: Addressed code review comments.