Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

DEN-15844: Fix to DEN-15865.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

  1. … 2 more files in changeset.
DEN-15844: Addressed DEN-15865 re: occlusion baseline not getting set.

  1. … 4 more files in changeset.
Done

Done

DENBUG-64 addressed the code review comments

  1. … 2 more files in changeset.
DENBUG-64 changed the code to constantly reset the blood sitting timer in treatment stop recover blood detections state

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DENBUG-64 updated the blood leak sensor code

  1. … 1 more file in changeset.
Done

Done

DENBUG-64 addressed the code review comments

  1. … 2 more files in changeset.
Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Should be the recovery alarm.

Should be the recovery alarm.

DENBUG-64 work in progress updated the blood detect algorithm

  1. … 1 more file in changeset.
DENBUG-64 work in progress with modifying the blood detection method

  1. … 3 more files in changeset.
Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

  1. … 1 more file in changeset.
DEN-15844: Addressed DEN-15710 re: state change event when starting rinseback mode.

HD-DEN-15844_Vbt Remains Open During Pressure Dry Self Test Triggering Alarm 23
HD-DEN-15844_Vbt Remains Open During Pressure Dry Self Test Triggering Alarm 23
Merged staging

  1. … 2 more files in changeset.