Merge remote-tracking branch 'origin/DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed' into develop

DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

cleared alarm 120 in self test

HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
Merge remote-tracking branch 'origin/DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed' into develop

  1. … 4 more files in changeset.
DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

Cleared alarm 120, prior to initiating alarm 85

Required latch closed before allowing alarm 85

Paused in WET self test until bubbles clear or latch closed.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into staging

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15664-blood-pump-stop-time-exceeded-alarms-124-125-165-and-212-alarm-timer-feature-is' into staging

Fixed. Thank you!

Fixed. Thank you!

Merge branch 'DEN-15664-blood-pump-stop-time-exceeded-alarms-124-125-165-and-212-alarm-timer-feature-is' into develop

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( ) This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysa...

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( )

This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysate flow everywhere, but is somewhat distributed.

Where is the heater turned back on? Is the state saved?

Where is the heater turned back on? Is the state saved?

Is this intended to be a BITWISE OR? Did you intend || ?

Is this intended to be a BITWISE OR? Did you intend || ?

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into staging

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into develop

DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus

Stop the pumps more forcefully, open loop, prior to switch valves for the saline bolus.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed overrides.

Removed overrides.

DEN-15664 Remove treatment end publish overrides.