Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15858 removed the generic messages timeout function

HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
DEN-15886: Addressed latest issue from DEN-15898 re: allowing bubble alarm during treatment recirculation state.

HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
Merge remote-tracking branch 'origin/DEN-16016-bug-unable-to-start-the-saline-bolus-3' into develop

Test build with open loop pump stop for saline bolus.

DEN-16016-bug-unable-to-start-the-saline-bolus-3

Re-reinstating earlier bolus change. DEN_15839

HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
Merge branch 'DEN-15512-dg-hd-failing-to-set-calibration-record' into develop

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-15716-skipped-treatment-start-state-while-transitioning-from-pre-treatment-to-treatment

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-15716-skipped-treatment-start-state-while-transitioning-from-pre-treatment-to-treatment

This is a second, separate branch addressing this ticket, and these changes should only be in develop (for testing) at this point. Eventually, this branch should replace the branch you reverted out...

This is a second, separate branch addressing this ticket, and these changes should only be in develop (for testing) at this point. Eventually, this branch should replace the branch you reverted out of staging.

stopDialysis() almost exactly mirrors the four lines I commented out, except the pump "hardstop" routines also manage the control variables that start and stop the PI controllers. This should still make zero difference to the case Nick is investigating since these changes are local to a saline bolus, which is not run at all while active ultra filtration is being run, and only affect the initial bolus state (0), while stopping the pumps. State 1, puts the pumps back into closed loop control, once the valves separate the patient, the blood and the dialysate from the dialyzer.

These changes are in staging right now. Do we have to stop dialysis in this state?

These changes are in staging right now. Do we have to stop dialysis in this state?

DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus-2

Reverted saline bolus change, updated to the wrapper function.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
DEN-15804 reverted the saline bolus changes