DEN-14815: Added support for post-tx next button request handling with acceptance and reject reason depending on whether reservoirs are drained and valves homed yet.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-59: Addressed some compiler warnings.

  1. … 3 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16102: code review comments

Fixed. Thanks!

Fixed. Thanks!

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged staging to DEN-15741.

DEN-16102: fixed operation status event to be called after current mode is updating

Remove this comment.

Remove this comment.

DEN-16102: changed dry self start state to always handle alarms with stop property.
DEN-16102: changed dry self start state to always handle alarms with stop property.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

DEN-15994-rinseback-5-minute-timer-does-not-count-down

Address CR comments.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Typo.

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not imple...

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not implement the mechanism to stop the pumps.

Implementing the change.
Done.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Blood pump only stops for user stop.

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circui...

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circuit (because rinseback was completed before we got to recirc), the NoBloodRecirc alarm property maybe doesn't apply here and we should only stop BP if the user stop alarm is active.

Merge remote-tracking branch 'origin/DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2' into develop

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Remove redundant condition.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.