Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-16175: moved end treatment timeout alarm above the substate execution.

DEN-16175: Moved End Treatment timeout alarm from wait for rinseback substate to general treatment end.

HD-DEN-16018_FW UF Control During Blood Leak Alarm Recover
HD-DEN-16018_FW UF Control During Blood Leak Alarm Recover
Remove commented / unused code. If it is to be used, it does not check for overflow.

Remove commented / unused code. If it is to be used, it does not check for overflow.

DEN-16018 added dialin and dialout pumps hard stop commands prior to transitioning to treatment from treatment pause

HD-DEN-15980_Pressure Alarm Windows Widened For Extended Period OF Time Due TO Frequency OF Air
HD-DEN-15980_Pressure Alarm Windows Widened For Extended Period OF Time Due TO Frequency OF Air
Add currentPostTreatmentState to header

Add currentPostTreatmentState to header

HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
DGFault instead of Dfault

DGFault instead of Dfault

DEN-14815: Added support for post-tx next button request handling with acceptance and reject reason depending on whether reservoirs are drained and valves homed yet.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-59: Addressed some compiler warnings.

  1. … 3 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16102: code review comments

Fixed. Thanks!

Fixed. Thanks!

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged staging to DEN-15741.

DEN-16102: fixed operation status event to be called after current mode is updating

Remove this comment.

Remove this comment.

DEN-16102: changed dry self start state to always handle alarms with stop property.
DEN-16102: changed dry self start state to always handle alarms with stop property.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

DEN-15994-rinseback-5-minute-timer-does-not-count-down

Address CR comments.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Typo.