Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-4211: Address code review feedback. Added additional UI message handling for treatment parameters.

  1. … 7 more files in changeset.
DEN-4211: Created saline bolus states.

  1. … 4 more files in changeset.
DEN-4211: Addressed code review feedback.

  1. … 11 more files in changeset.
DEN-4211: Changed treatment parameter order in messaging per agreement with Peter.

DEN-4211: Merge with doxygen fix branch.

  1. … 2 more files in changeset.
DEN-4534: More doxygen udpates.

  1. … 9 more files in changeset.
DEN-4534: Doxygen updates.

  1. … 30 more files in changeset.
DEN-4211: Added stop treatment logic when treatment duration is reached.

  1. … 2 more files in changeset.
DEN-4308 added a temporary struct to publish the valves data every 10ms for characterization

  1. … 4 more files in changeset.
DEN-4211: Added treatment parameter override message.

  1. … 3 more files in changeset.
DEN-4211: Added more treatment parameter validation logic and messaging.

  1. … 6 more files in changeset.
DEN-4308 Added comments and headers to the functions in valves.c

  1. … 6 more files in changeset.
DEN-4308 worked on the valves homing sequence

  1. … 6 more files in changeset.
DEN-4211: Updated format of @params in all modules.

  1. … 21 more files in changeset.
DEN-4211: Fixed Vectorcast math.h issue and removed band-aid macros in common.h.

  1. … 5 more files in changeset.
DEN-4211: Added 2 sec delay before power off. Added HD valves support to EMC build.

  1. … 4 more files in changeset.
DEN-4211: Finished HD pumps open loop Dialin command update. Started data infrustructure for treatment parameters mode.

  1. … 13 more files in changeset.
DEN-4212: Tuned blood and dialysate flow sensors. Changed blood and dialysate pump control back to closed loop. Updated FPGA ID.

  1. … 5 more files in changeset.
DEN-4212, DEN-4217: Added accelerometer feature. Added calibration to blood and dialysate flow. Moved msg payload structures to a separate header file.

  1. … 13 more files in changeset.
Matched DG arrangement and fix for CAN Tx. Added inlet conductivity alarm ID.

  1. … 5 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 21 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 56 more files in changeset.
UT updates. Added UT for DGInterface. Misc fixes per UT.

  1. … 4 more files in changeset.
Changed UF reference volume update approach from time x rate equation to accumulation of step time x rate. Fixed debug logging of event data when alarm is triggered. Updated flow test build output.

  1. … 4 more files in changeset.
Added misc test/debug build flags. Added persistence to CAN buffer overflow faults so that we can try to recover from one-off spurts when a node is coming online.

  1. … 8 more files in changeset.
Fixed UF change verification logic. Fixed doxygen comments.

  1. … 3 more files in changeset.
Made reservoir target volumes public. Set all UF volumes to F32 type for consistency.

  1. … 4 more files in changeset.
Addressed code review feedback. Fixed treatment time minimum range to UI message.

  1. … 1 more file in changeset.
Fixed message payload sizing issue on UF option confirmation message handling.

  1. … 1 more file in changeset.
Changed max dialysate volume from 180 to 150 L. Split UF change response messages into 2 separate messages to prevent UI confusion.

  1. … 2 more files in changeset.