DEN-14561 Updated reservoir switching

Moved active reservoir marker into local states, reinserted the MGMT complete state change and the SW_CONFIG test volumes

DEN-14561: fixes to pre-tx fill cycles.

DEN-14561

Reservoir selection update.

DEN-14549: Added alarm threshold to occlusion alarm data for logging. Only check for sufficient Heparin when in pre-tx mode. Addressed code review comments.

  1. … 2 more files in changeset.
DEN-14561

Partial fix for fill volumes pretreatment

Merge branch 'DEN-14718-dry-self-test-failure' into develop

DEN-14718: Increased air pump to 60s timeout. Fixed air pump not turning off when alarm 231 is triggered in self test state.

DEN-14549: Addressed bugs DEN-13769 and DEN-14231.

DEN-14561 Corrected air pump state.

After a dry self test failure, turned off the air pump.

DEN-14549: Fixed build issue.

DEN-14561-_release_-preprocessor-flag-specifies-code-that-conflicts-with-sw_config_-debug into develop

Merge branch

  1. … 2 more files in changeset.
DEN-14549: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-14549: Added fpga support for syringe pump DAC registers. Addressed code review comments.

  1. … 3 more files in changeset.
HD-DEN-14561

Simplified and corrected reservoir management for multiple cycle flush

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14589-DEN-14011_fix' into develop

  1. … 1 more file in changeset.
DEN-14011: Fixed hanging dry self tests.

HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug

Corrected DISABLE_AIR_PUMP logic, to allow reservoir fill to proceed when the air pump is disabled.

DEN-14549: Update to arterial/venous pressure limit alarm logic. Update measured UF volume in all dialysis states. Remove dialysis and UF start states.

  1. … 4 more files in changeset.
HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug

Incorporated code review observations.

DEN-14561 _RELEASE_ preprocessor flag specifies code that conflicts with SW_CONFIG_ debug flags

Require one initial drain.

Simplify.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-14011-occulsion-alarm-when-reseating

DEN-14549: Fix art/ven pressure alarm check logic - apply user set limits in dialysis state (if not saline bolus) or treatment stop state (if BP running). Require door/latch to be closed in patient connection state. Remove escalation for prime complete alarms. Handle prime complete alarms in pre-tx recirculate state. Update prime complete alarm properties.

  1. … 2 more files in changeset.
DEN-14011

DEN-14561 Wait for dry self test before final fill. Use reservoir 2 for Prime dialysate bypass.

DEN-14011: implemented proper exit condition for dry self start state.

DEN-14011: created handleDrySelfTestStartState. code review comments.

  1. … 1 more file in changeset.
DEN-14561 Update PRE_TREATMENT_FILL_RESERVOIR_TWO_VOLUME_ML per code review comment.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces