AlarmMgmt.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-13865: fix re-trigger block logic.

DEN-13865: Handle no re-trigger action or end-tx only alarms during rinseback/recirc states of treatment mode.

  1. … 2 more files in changeset.
DEN-13865: Implemented alarm system feature to prevent alarm re-trigger per new properties. Addressed DEN-15295.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15060-2: Another fix to alarm silence functionality.

DEN-15060-2: Fixed issue with alarm silence getting aborted by a lesser ranked alarm.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'staging' into DEN-15060-nfr-alarms-lower-priority-alarm-de-silence-higher-priority-alarm-audio

DEN-15060: Addressed DEN-15060 issue re: new alarms ending alarm silence.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15245: Addressed DEN-15245.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
DEN-14262: Added a/c power loss flag to alarm info broadcast.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Merge branch 'DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming' into staging

Merge branch 'DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming' into develop

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Addressed code review comments.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Addressed paren typo from code review

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Additional Code review comments addressed.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Moved Alarms blocked timer variable, per review.

  1. … 1 more file in changeset.
DEN-15183: Fixed compiler warning.

DEN-15183: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Comments updated.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Updated allowed alarms during PF

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Switched to CPLD power loss function, rather than alarm.

DEN-15183: Prevent alarm user options while AC power is out.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Moved block to choke point in activateAlarm2Data. Bock all alarms, rather than sub-rank 10 and above.

DEN-15183: Implemented updates to treatment recirculation workflow to include states for instructing user to disconnect and reconnected patient. Updated scalar for alarm audio current high gain signal. Updated alarm audio POST.

  1. … 2 more files in changeset.
DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit