DEN-4640: Added alarm for weak flow sensor signal strength. Added handling of saline bolus abort from user.

  1. … 3 more files in changeset.
DEN-4308 added another define for a larger range check in Position B during current relaxation

  1. … 1 more file in changeset.
DEN-5053: UT updates.

  1. … 1 more file in changeset.
DEN-4308 merged master, resolved merge conflicts and continues development testing

  1. … 16 more files in changeset.
DEN-4308 addressed code review comments and continued dev testing

  1. … 4 more files in changeset.
DEN-4640: Further developed saline bolus functionality.

  1. … 12 more files in changeset.
DEN-5053: Added air trap alarms. Added actuator safe state enforcement to fault mode.

  1. … 7 more files in changeset.
Added new HD valve alarm to table.

DEN-5053: Fixed doxygen function headers for inputs/outputs details section.

  1. … 36 more files in changeset.
DEN-4308 Addressed code review comments and continued development testing

  1. … 6 more files in changeset.
DEN-5053: Addressed code review feedback. Fixes to treatment workflow messaging with UI. Added an alarm lamp sync flag to alarm status message and increased transmission rate from 2 Hz to 4 Hz to allow alarm lamp sync flag to reach UI in timely manner. Added an assertion to give compiler error if new alarm is added to enumeration and not to alarm definition table. Update FPGA UT.

  1. … 22 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 60 more files in changeset.
DEN-4211: Updated unit tests for treatment param mode and utilities modules. Added range and range checking to critical data feature in utilities module. Updated UF rate accuracy check.

  1. … 3 more files in changeset.
DEN-4211: Addressed code review feedback.

  1. … 15 more files in changeset.
DEN-4534: More doxygen updates.

  1. … 12 more files in changeset.
DEN-4534: Doxygen updates.

  1. … 38 more files in changeset.
DEN-4211: Added more treatment parameter validation logic and messaging.

  1. … 10 more files in changeset.
DEN-4211: Updated format of @params in all modules.

  1. … 30 more files in changeset.
DEN-4211: Updated IT for Utilities module. Updated UT/IT for NVData module.

DEN-4217: Added accelerometer alarms.

DEN-4211: Finished HD pumps open loop Dialin command update. Started data infrustructure for treatment parameters mode.

  1. … 15 more files in changeset.
DEN-4184: Added new alarms to HD alarms

DEN-4169: Removed extra line at the end

DEN-4169: Updated HD alarms

Matched DG arrangement and fix for CAN Tx. Added inlet conductivity alarm ID.

  1. … 5 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 24 more files in changeset.
Added f/w debug message support. Added debug messages for CAN/UART error interrupts and alarm trigger/clear events. Added FPGA version info to version message. Added handler for DG/UI alarm trigger/clear events. Added debug test code for task/interrupt timing.

  1. … 16 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 70 more files in changeset.
Changed UF reference volume update approach from time x rate equation to accumulation of step time x rate. Fixed debug logging of event data when alarm is triggered. Updated flow test build output.

  1. … 4 more files in changeset.
Updated FPGA register map to latest HDD. Separated ramp up and ramp down rates so they can be set differently for blood and dialysate pumps.

  1. … 6 more files in changeset.