I think it should be and. Because it is a "whichever is greater" condition. So if error is 6% but delta is less than 0.1 mL, I would not want to trigger the error because the volume is too small to...
I think it should be and. Because it is a "whichever is greater" condition. So if error is 6% but delta is less than 0.1 mL, I would not want to trigger the error because the volume is too small to alarm on a percentage basis. Likewise, if delta is > 0.1 mL but error is < 5%, no alarm.