Late breaking news, the HD_PRE_TREATMENT_PRIME_STATE_T has changed. The correct order of the items in enum type of the pre-treatment states are listed below: HD_PRIME_START_STATE = 0, ///< Prime s...
Late breaking news, the HD_PRE_TREATMENT_PRIME_STATE_T has changed.
The correct order of the items in enum type of the pre-treatment states are listed below:
HD_PRIME_START_STATE = 0, ///< Prime start state
HD_PRIME_WAIT_FOR_USER_START_STATE, ///< Wait for user to start prime state
HD_PRIME_SALINE_SETUP_STATE, ///< Saline setup state
HD_PRIME_SALINE_PURGE_AIR_STATE, ///< Saline purge air state
HD_PRIME_SALINE_CIRC_BLOOD_CIRCUIT_STATE, ///< Circulate blood circuit state
HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE, ///< Wait for reservoir 1 fill complete
HD_PRIME_DIALYSATE_DIALYZER_STATE, ///< Dialysate dialyzer fluid path state
HD_PRIME_SALINE_DIALYZER_SETUP_STATE, ///< Saline dialyzer setup state
HD_PRIME_SALINE_DIALYZER_STATE, ///< Saline dialyzer fluid path state
HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE, ///< Wait for reservoir 2 fill complete
HD_PRIME_DIALYSATE_BYPASS_STATE, ///< Dialysate bypass fluid path state
HD_PRIME_WET_SELF_TESTS_STATE, ///< Perform wet self-tests after priming complete
HD_PRIME_PAUSE, ///< Prime pause state, waits to be resumed
HD_PRIME_COMPLETE, ///< Prime complete state
NUM_OF_HD_PRIME_STATES ///< Number of prime sub-mode states