Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merged DEN-14175

  1. … 6 more files in changeset.
DEN-14175 replaced the V3 flag with FPGA hardware check and added the default the fixed valves PWM count in FPGA

  1. … 1 more file in changeset.
0.0F

0.0F

0.0F

0.0F

0.0F

0.0F

DEN-14170 Fix TODO with new fault.

  1. … 1 more file in changeset.
HD-DEN-14170_SW Dev Sprint 82 Darren
HD-DEN-14170_SW Dev Sprint 82 Darren
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Update function header.

Update function header.

Remove blank line

Remove blank line

Done.

Done.

This function should stay commented. I have a TODO task for this.

This function should stay commented. I have a TODO task for this.

Done.

Done.

No these should be TRUE.

No these should be TRUE.

DEN-14175 updated the reservoir switch command to send the trimmer use the previous duty cycle or estimate the duty cycle

  1. … 7 more files in changeset.
Add a space between "return" and "(".

Add a space between "return" and "(".

Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop

  1. … 2 more files in changeset.
DEN-14175 implemented the FGPA hardware configuration register, changed the trimmer heater start/stop command to HD command and not the log in command. Added a command to benign the polynomial calibration values of the temperature sensors

  1. … 4 more files in changeset.
hasReg1Failed is initialized to FALSE. Should we be setting to TRUE in these ifs? Or should the initialize be setting to TRUE?

hasReg1Failed is initialized to FALSE. Should we be setting to TRUE in these ifs? Or should the initialize be setting to TRUE?

This revision also applies to UI.

This revision also applies to UI.

Remove

Remove

Remove TODO comment.

Remove TODO comment.

Merge branch 'DEN-14150-hd-dialysate-flow-estimation' into develop

  1. … 2 more files in changeset.
DEN-14150: Fixes to DPi rotor count override.

  1. … 2 more files in changeset.
DEN-14150: Reset rotor counts when starting new treatment workflow. Added DPi rotor count override message handling.

  1. … 3 more files in changeset.
HD-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
HD-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
Make #define for 0.5.

Make #define for 0.5.

Should use get function instead of filtered.... We would want override to affect this check. Probably means we have to move this check to function that calls this function after override structure ...

Should use get function instead of filtered.... We would want override to affect this check. Probably means we have to move this check to function that calls this function after override structure is set to new estimate.

Please update the DIP_ML_PER_MIN_FROM_PWM to reflect the changes in DIP_PWM_FROM_ML_PER_MIN. DIP_ML_PER_MIN_FROM_PWM should be the inverse equation of DIP_PWM_FROM_ML_PER_MIN. The rate in ml/min as...

Please update the DIP_ML_PER_MIN_FROM_PWM to reflect the changes in DIP_PWM_FROM_ML_PER_MIN. DIP_ML_PER_MIN_FROM_PWM should be the inverse equation of DIP_PWM_FROM_ML_PER_MIN. The rate in ml/min as an output with pwm as its input.