DEN-14549: Misc updates to dry self-tests in pre-treatment.

  1. … 2 more files in changeset.
HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
DEN-14549: Fixed issue where door open alarm would have condition cleared repeatedly in standby start state. Made top alarm determination done from scratch each time.

  1. … 1 more file in changeset.
DEN-14549: Removed obsolete internal ADC channel. Updates internal ADC channel transfer functions. Misc pre-tx state optimizations.

  1. … 6 more files in changeset.
It is in the header under Outputs.

It is in the header under Outputs.

Add dgServiceAndUsageData to header.

Add dgServiceAndUsageData to header.

DEN-14549: Updates to reasons to reject start treatment request at standby mode.

  1. … 5 more files in changeset.
HD-DEN-14559_DG HD Dev HD DG Dvt Update Part 8
HD-DEN-14559_DG HD Dev HD DG Dvt Update Part 8
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14497 addressed the code review comments

Done

Done

There are many of these - not just this one.

There are many of these - not just this one.

DEN-14497 addressed the code review comment

Done

Done

Why are these changed back to ACK_NOT_REQUIRED?

Why are these changed back to ACK_NOT_REQUIRED?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'staging' into DEN-14497-dg_hd_dev-hd_dg_dvt-update-part-7

  1. … 5 more files in changeset.
DEN-14497 addressed the code review comments

  1. … 1 more file in changeset.
Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.

Fixed merge conflicts.

  1. … 15 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.