They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing...

They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing and is part of treatment mode module.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Done.

Done.

Done.

Done.

Done.

Done.

Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
It is in the header under Outputs.

It is in the header under Outputs.

Add dgServiceAndUsageData to header.

Add dgServiceAndUsageData to header.

HD-DEN-14559_DG HD Dev HD DG Dvt Update Part 8
HD-DEN-14559_DG HD Dev HD DG Dvt Update Part 8
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

There are many of these - not just this one.

There are many of these - not just this one.

Done

Done

Why are these changed back to ACK_NOT_REQUIRED?

Why are these changed back to ACK_NOT_REQUIRED?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

FALSE was correct (throughout) for last param of sendAckResponseMsg() if message/request found to be invalid in some way, TRUE if message is ok.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.

Timer variable is actually a time stamp of last command being queued - variable name could be clearer.