Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15375 addresed the release fixes

HD-DEN-15375_FW Dialysate Heating Fixes
HD-DEN-15375_FW Dialysate Heating Fixes
HD-DEN-15250_Pre Treatment Does Not End Successfully
HD-DEN-15250_Pre Treatment Does Not End Successfully
This doesn't seem like a fix to root cause - may not work if user clears alarm quickly. Suggest looking at valve driver where alarm 62 is detected/triggered and maybe zeroing positionOutOfRangeCoun...

This doesn't seem like a fix to root cause - may not work if user clears alarm quickly.
Suggest looking at valve driver where alarm 62 is detected/triggered and maybe zeroing positionOutOfRangeCounter (reset 1 sec persistence) if AC power is out. That way, when AC is restored, valves will have a whole second to get where they're supposed to be before alarm is triggered.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15060-2: Another fix to alarm silence functionality.

HD-DEN-15060_2 Nfr Alarms Lower Priority Alarm DE Silence Higher Priority Alarm Audio
HD-DEN-15060_2 Nfr Alarms Lower Priority Alarm DE Silence Higher Priority Alarm Audio
DEN-15060-2: Fixed issue with alarm silence getting aborted by a lesser ranked alarm.

Merged staging

  1. … 1 more file in changeset.
HD-DEN-15367_HD DG Test Configuration Setup
HD-DEN-15367_HD DG Test Configuration Setup
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
Done

Done

Add patientConnectionConfirm and doorOpened to header.

Add patientConnectionConfirm and doorOpened to header.

Merge branch 'DEN-15274-hd_dg_test_configuration_setup' into develop

DEN-15274, DEN-15360 changed the code to not alarm watchdog expired until POST is done

HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power
HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power
When will commented code be removed.

When will commented code be removed.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done

Merged DEN-15274

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.