DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Corrected alarm block condition

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into develop

  1. … 1 more file in changeset.
Merge branch 'staging' into DENBUG-64-sys-dialyzer-blood-leak-alarm-triggered-during-treatment

  1. … 2 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed merge conflict.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done

Done

Done.

Done.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

  1. … 2 more files in changeset.
DEN-15844: Addressed DEN-15865 re: occlusion baseline not getting set.

  1. … 3 more files in changeset.
Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Should be the recovery alarm.

Should be the recovery alarm.