Services

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-6078 addressed some of the merge issues

  1. … 3 more files in changeset.
DEN-7091: Added Sample Water module

  1. … 6 more files in changeset.
DEN-6372: Moved blood sitting timeout logic from alarm escalation to treatment stop sub-mode.

  1. … 14 more files in changeset.
DEN-5980 disbaled some of the execs to be able to test NVDataMgmt

  1. … 6 more files in changeset.
DEN-5887: Fixed merge conflicts from staging.

  1. … 10 more files in changeset.
DEN-6372: Fixes per lab testing. Added function to AlarmMgmt to determine whether any alarms w/ stop property are active. Changed treatment mode handling of alarm stop from signal to modes approach to modes polling alarm mgmt approach. Added f/w support for override of treatment time remaining.

  1. … 3 more files in changeset.
DEN-6078 merged staging and resolved merge conflicts

  1. … 2 more files in changeset.
DEN-6372: Addressed code review feedback.

  1. … 1 more file in changeset.
DEN-6372: Fixed bug in treatment recirc progress message broadcast.

DEN-5980 removed old calibration functions from the drivers

  1. … 2 more files in changeset.
DEN-5980 added more NV comm messages

  1. … 1 more file in changeset.
DEN-5980 added comm messages to be able to set and get NV records

  1. … 11 more files in changeset.
DEN-6372: Added timeout & countdown to rinseback and recirc status messages to UI.

  1. … 3 more files in changeset.
DEN-6402: Added DG valve setting change command

  1. … 5 more files in changeset.
DEN-6402: Stored command response for each possible command

  1. … 3 more files in changeset.
DEN-5887: Refactored load cell acceleration alarm. Tuned UF control.

  1. … 3 more files in changeset.
DEN-6372: consolidated dialysis transition and start functions. Re-initialize treatment modes before starting a new treatment workflow.

  1. … 5 more files in changeset.
DEN-6402: Added alarm signal action handlers

  1. … 7 more files in changeset.
DEN-6372: Amended treatment stop sub-mode.

  1. … 5 more files in changeset.
DEN-6372: Added treatment stop state to treatment states message.

  1. … 4 more files in changeset.
DEN-6372: Updated treatment state broadcast msg to include blood prime and treatment end sub-mode states.

  1. … 3 more files in changeset.
DEN-6372: Added new blood prime sub-mode of treatment mode.

  1. … 15 more files in changeset.
DEN-6372: Added user messaging for treatment end sub-mode.

  1. … 1 more file in changeset.
DEN-5887: Increase load cell change threshold for non-treatment modes so will not cause alarm when load cell is tared. Remove obsolete messaging.

  1. … 3 more files in changeset.
DEN-6402: Removed home setup state and refactor big functions

  1. … 7 more files in changeset.
DEN-6402: Removed user end treatment request

  1. … 1 more file in changeset.
DEN-6402: Added blood leak no cartridge self-test

  1. … 1 more file in changeset.
DEN-6372: Minor fixes. Added publishing of rinseback status to UI.

  1. … 5 more files in changeset.
DEN-6402: Added wet self-tests state machine

  1. … 5 more files in changeset.
DEN-6402: Added air bubble detector self-tests

  1. … 2 more files in changeset.