Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5674: Make all code comments start with a capital letter for consistency.

  1. … 45 more files in changeset.
DEN-5674: Addressed code review comments.

  1. … 21 more files in changeset.
DEN-5674: Added msg ID for auto-ACK to enum. Misc. comments. Removed dead code.

  1. … 1 more file in changeset.
DEN-5674: Changed size of alarm silence cmd from U08 to U32 for consistency.

DEN-5381: Reset pre-treatment state machine and deleted unused code

  1. … 2 more files in changeset.
DEN-5674: Fixed alarm audio volume control per HDD change.

DEN-5381: Moved new alarms to available slots

DEN-5381: Updated hd to dg drain cmd

  1. … 3 more files in changeset.
DEN-5381: Removed reservoir drain and fill from standby mode

  1. … 3 more files in changeset.
DEN-5674: Added handling for resume after alarm. Removed obsolete build switches.

  1. … 15 more files in changeset.
DEN-5674: Fix alarm state bug.

DEN-5674: Added alarm silence handling.

DEN-5674: temporary set alarm volume low for testing.

DEN-5674: Fixed alarms state update.

DEN-5674: Prevent resume/rinseback options if top alarm condition not yet cleared.

  1. … 1 more file in changeset.
DEN-5381: Added reservoir management to prime sub-mode

  1. … 2 more files in changeset.
DEN-5674: Removed obsolete test code. Added new UF param msg handler.

  1. … 14 more files in changeset.
DEN-5381: Refactor alarm management module

DEN-5674: Fixed valves POST call. Added alarm for FPGA comm timeout.

  1. … 1 more file in changeset.
DEN-5674: Removed alarm dialyzer bypass flag. Added alarm user option for alarms that simply require acknowledgement. Added some alarm action handling for treatment mode.

  1. … 6 more files in changeset.
DEN-5381: Created skeleton for pre-treatment mode

  1. … 3 more files in changeset.
DEN-5674: Added mode specific masking of enabled user alarm recovery actions.

  1. … 8 more files in changeset.
DEN-5674: Fixed drain reservoir cmd to DG to include indication of whether reservoir load cells should be tared.

  1. … 3 more files in changeset.
DEN-5674: Added alarm (empty) action signal handling functions to the various op modes.

  1. … 18 more files in changeset.
DEN-5674: Added alarm sub-rank property and use for determining top priority alarm for UI.

  1. … 1 more file in changeset.
DEN-5674: Added synchronize of alarm audio/lamp control with published alarm state so that UI can flash alarm in sync w/ lamp. Added alarm audio.

  1. … 2 more files in changeset.
DEN-5674: Updated alarm messaging. Added concept of alarm condition detected and ability to clear this. Clear all recoverable alarms when user chooses an action.

DEN-5887: Fixed name format for const data to all caps w/ underlines per coding standard.

  1. … 5 more files in changeset.
DEN-5887: Added controller type property to PI controller as was added in DG stack. Added max I step size to prevent lasting effects of a large but brief disturbance in control input.

DEN-5887: Moved some load cell functions from dialysate outlet pump controller to DG Interface module. Added small and large moving average filters to load cell readings. Use small moving average for normal UF tracking/control and use large moving average for baseline and final reservoir weights to minimize reservoir transition errors due to load cell noise.

  1. … 6 more files in changeset.