Services

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14999 updated the alarms table

  1. … 20 more files in changeset.
Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

DEN-14830: fixed alarm mgmt faulting on clear.

Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 1 more file in changeset.
DEN-14830: Fixed alarm mgmt not clearing alarm ids under 30. Fixed Dial temps

  1. … 1 more file in changeset.
Merge branch 'DEN-14999-alarm-table-updates-part-2' into develop

  1. … 2 more files in changeset.
DEN-14999-2: Fixed bug in UF rate alarm detection.

DEN-14999-2: Removed alarm escalation feature.

  1. … 1 more file in changeset.
Merge branch 'DEN-14999-alarm-table-updates-part-2' into develop

  1. … 3 more files in changeset.
DEN-14999-2: prevent treatment resumption from rinseback/recirc states if an alarm that blocks resume has been triggered previously in the treatment.

  1. … 4 more files in changeset.
DEN-14999 removed the depletion alarm (alarm 168)

Merged staging

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 17 more files in changeset.
DEN-14999: merged staging

  1. … 2 more files in changeset.
Merge DEN-15016 into develop

DEN-14999: Additional fix to DEN-14850.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Merge DEN-14860 to staging

  1. … 1 more file in changeset.
HD-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen

Closing code review comments.

DEN-14999: Small fix to HD WD POST.

Merge branch 'DEN-14999-alarm-table-updates' into develop

  1. … 1 more file in changeset.
DEN-14999, DEN-15030 fixed the start DG command to only get the acid and bicarb parameters if the HD is sending a start command

Merge branch 'DEN-14999-alarm-table-updates' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-14999-alarm-table-updates

DEN-14999: Fixed sending of DG valve settings command.

DEN-14999, DEN-15030 updated the DG start command payload

Merge DEN-14860 branch

DEN-14999: Removed obsolete/redundant alarms. Addressed code review comment.

  1. … 2 more files in changeset.
DEN-15016-system-faults-during-treatment-then-returns-to-the-standby-screen

Increased message ACK timeouts, for both DG and UI.