DEN-13865: More fixes to alarm re-trigger prevention logic.

DEN-13865: fix re-trigger block logic.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-13865: Handle no re-trigger action or end-tx only alarms during rinseback/recirc states of treatment mode.

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

DEN-13865: Implemented alarm system feature to prevent alarm re-trigger per new properties. Addressed DEN-15295.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15367 updated the use wet cartrdige test configuration

So then can we remove power loss alarm active checks from this condition (just keep the cpld power loss check as before)?

So then can we remove power loss alarm active checks from this condition (just keep the cpld power loss check as before)?

Do we need the alarm active checks? Isn't power loss check sufficient?

Do we need the alarm active checks? Isn't power loss check sufficient?

HD-DEN-13865_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
HD-DEN-13865_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15250-pre-treatment-does-not-end-successfully' into develop

DEN-15250: cosmetic and coding standard changes.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge remote-tracking branch 'origin/DEN-15363-fw-handle-bld-sensor-communication-errors' into develop

DEN-15363 Reuse alarm ID 63 for Blood Leak communications errors

The root cause is that abnormal AC power loss on the HD causes abnormal operating conditions and spurious errors. The valves return to position after the alarm is cleared, not when AC returns. The ...

The root cause is that abnormal AC power loss on the HD causes abnormal operating conditions and spurious errors. The valves return to position after the alarm is cleared, not when AC returns. The timer gives a ten-second window following removing the power fail signal and the alarm condition, where the alarm block is active. The original block used the alarms alone, we changed to the CPLD power loss detection - which returns with the return of AC. Clearing the alarms late means removing the block before the valves return to normal operation.

I think there are two approaches. We already have distributed failsafe that is entered on power fail, which we recover from after the alarm is cleared.
1. Create a pervasive state that affects every operating mode, allowing transitions to "safe" states during alarms, then back to the original operating state. The safe state would not register errors until after the transition back to normal operation.
2. Block errors until both AC is returned and the alarm is cleared when normal operation is restored.

The second approach is much simpler and addresses the root problem: odd control configurations resulting from having alarms with conflicting controls. In this case, the secondary alarms are spurious since they result from the valves dropping into a safe position, for instance, on the loss of AC. We also have a handful of other errors that occur under these conditions, fan speed, for instance, as well as valve positions,

Added alarm 62 detection disable as suggested.

Added alarm 62 detection disable as suggested.

HD-DEN-15396_2 Alarm 42 HD Fault Dpo Rotor Encountered IN Blood Prime Dvt 00
HD-DEN-15396_2 Alarm 42 HD Fault Dpo Rotor Encountered IN Blood Prime Dvt 00
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflict.

Fixed merge conflict.

DEN-15250: Addressed DEN-15308.

DEN-15396-2: Fixed rotor speed chk persistence function to handle negative motor speeds.

DEN-15250: End tx should go to post-tx from pre-tx unless haven't finished water sample state. Require ack on confirmation msgs.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces