Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DENBUG-64 changed the blood detect code to not check for blood when the dialysate line is in bypass mode

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

DEN-15844: Fix to DEN-15865.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'DEN-15844-vbt-remains-open-during-pressure-dry-self-test-triggering-alarm-231' into develop

DEN-15844: Addressed DEN-15865 re: occlusion baseline not getting set.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-64-sys-dialyzer-blood-leak-alarm-triggered-during-treatment' into develop

Done

Done

DENBUG-64 addressed the code review comments

DENBUG-64 changed the code to constantly reset the blood sitting timer in treatment stop recover blood detections state

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-64-sys-dialyzer-blood-leak-alarm-triggered-during-treatment' into develop

DENBUG-64 updated the blood leak sensor code

Done

Done

DENBUG-64 addressed the code review comments

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.