Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged staging and resolved the conflicts

DEN-15105: Removed alarm 227. Fixed bug DEN-15102.

DEN-15110 Use 800ml/min reservoir fill rate for all dialysate flow rates.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15101: disable ac power loss alarms until cpld change gets propagated.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
Merge branch 'staging' into DEN-14999-alarm-table-updates-part-2

DEN-14999-2: Restored a/c power loss alarm detection.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14999 updated the alarms table

  1. … 6 more files in changeset.
DEN-14999-2: Addressed code review comment.

Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

DEN-14830: fixed alarm mgmt faulting on clear.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-14999-2: Temporarily disable AC power loss alarm until CPLDs are updated in lab.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

DEN-14830: Fixed alarm mgmt not clearing alarm ids under 30. Fixed Dial temps

DEN-14830: fixes

DEN-14830: simplifying temperature control logic.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14999-alarm-table-updates-part-2' into develop

DEN-14999-2: Increased UF alarm thresholds to 100 mL for volume and 10 mL/min for rate.

DEN-14999-2: Fixed bug in UF rate alarm detection.

DEN-14999 updated the code based on the alarms table