Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled the blood leak alarms

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-11250 Fixed merge conflicts

    • -171
    • +180
    ./Controllers/SyringePump.c
DEN-11250 Updated per peer review comments

DEN-11250 Fixed compile errors

  1. … 1 more file in changeset.
DEN-11250 Updated handPOSTStateStart with same function from DEN-11750 branch

DEN-11250 Updated per peer review comments

DEN-11750 made changes for the reservoir test

DEN-11250 Changed checkOcclusionOr Empty() to checkSyringeEmpty()

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled prime

Changed the reservoirs management for testing

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled the software configurations

DEN-11250 Stopped syringe pump then transition to dry self tests stopped state when alarm is triggered

DEN-11250 Incorporated peer review comments

  1. … 2 more files in changeset.
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Disabled a software configuraion

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-11750 and resolved the merge conflicts

DEN-11750 updated the blood leak driver to send Ctrl U multiple times and make sure the sensor is communicating

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Changed the BLD self test timeout

DEN-11250 Merged branch DEN-11750 dg_dev-dg dialysate temperature tune up to branch DEN-11250 hd syringe pump update before working on syringe pump bug ticket #11824

    • -171
    • +188
    ./Controllers/SyringePump.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Enabled the build switches

DEN-11250 Implemented new requirements for syringe pump occlusion & empty detection - when prime is done, wait 1 sec, calc delta force sensor greater than 0.5then trigger occlusion alarm