DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

    • -0
    • +2
    /firmware/App/Modes/ConsumableSelfTest.c
DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

    • -0
    • +2
    /firmware/App/Modes/ConsumableSelfTest.c
DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 Fixed ticket DEN-12533 HD should accepting only value for arterial pressure high from -270 to 0

    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-12224 Added build switch to disable syringe pump removed alarm[219] on system without syringe pump installed

DEN-10981 Fixed options to resume/rinseback are present after removing heparin syringe

DEN11250 Added code to allow resume after syringe pump occlusion has been removed - moved here from commit 1bb17047684 in DEN-11250 branch

DEN-11250 Fixed syringe pump occlusion alarm not triggered - moved commit 269329748ff from DEN-11250 to here

    • -4
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11928 Addressed peer review comments

DEN-11928 Fixed compile errors

DEN-11928 Merged commit 2df3175ad36 to DEN-11928 branch

DEN-11928 Changed heparin bolus rate from 40.0 to 720.0 mL/hr per Systems request

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11928 Changed heparin bolus target rate from 40.0 to 720.0 mL/hr per Systems request

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11928 Added SW_CONFIG changes because committed 667feab319b 10Feb 2022 was missing in committed 3d72b777cf1 15Feb 2022

    • -8
    • +14
    /firmware/App/Modes/ConsumableSelfTest.c
DEN-11928 Updated code due to changes in DG_Fill_Mode_States in Common

    • -4
    • +4
    /firmware/App/Modes/ConsumableSelfTest.c
DEN-11928 Reverted conductivity changes

    • -17
    • +11
    /firmware/App/Modes/ConsumableSelfTest.c
DEN-11250 Added code to allow resume after syringe pump occlusion has been removed

DEN-11250 Added code to allow resume after the syringe pump occlusion has been removed

DEN-11250 Fixed syringe pump occlusion alarm not triggered

    • -4
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11250 Fixed syringe pump occlusion alarm not triggered

    • -4
    • +1
    /firmware/App/Controllers/SyringePump.c
Merge branch 'DEN-11250-hd-syringe-pump-updates' into develop

Merge branch 'DEN-11250-hd-syringe-pump-updates' into staging

DEN-11250 Installed case MSG_ID_HD_FAN_RPM_ALARM_START_TIME_OFFSET_OVERRIDE per peer review comments

DEN-11250 Fixed merge conflicts

    • -30
    • +12
    /firmware/App/Controllers/PresOccl.c
    • -97
    • +88
    /firmware/App/Controllers/SyringePump.c
DEN-11250 Fixed merge conflicts

    • -23
    • +7
    /firmware/App/Controllers/PresOccl.c
    • -183
    • +23
    /firmware/App/Services/SystemComm.c
DEN-11250 Init ovData field to 40.0 mL/hour

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11250 Updated per peer review comments

    • -108
    • +37
    /firmware/App/Controllers/SyringePump.c
    • -1
    • +0
    /firmware/App/Controllers/SyringePump.h
DEN-11250 Fixed function getHeprinBolusTargetRate declared implicitly

    • -0
    • +1
    /firmware/App/Controllers/SyringePump.h
DEN-11250 Fixed occlusion alarm did not triggered. ForceAtEndOfPrime = 0.6V with 1 second delay, ForceAtEndOfPrime = 2.5V with 3 seconds delay

DEN-11250 Do not trigger syringe pump running while BP is off alarm #83 when syringe pump is in heprin bolus state

    • -3
    • +1
    /firmware/App/Controllers/SyringePump.c