DEN-11250 Fixed merge conflicts

    • -23
    • +7
    /firmware/App/Controllers/PresOccl.c
    • -183
    • +23
    /firmware/App/Services/SystemComm.c
DEN-11250 Init ovData field to 40.0 mL/hour

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11250 Updated per peer review comments

    • -108
    • +37
    /firmware/App/Controllers/SyringePump.c
    • -1
    • +0
    /firmware/App/Controllers/SyringePump.h
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Commented out the alarms temporarily

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Removed DEBUG_DENALI

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

Initialized the variables

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-11750 and resolved the merge conflicts

    • -16
    • +19
    /firmware/App/Controllers/DGInterface.c
    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
    • -29
    • +80
    /firmware/App/Controllers/PresOccl.c
    • -1
    • +3
    /firmware/App/Controllers/SyringePump.c
    • -4
    • +12
    /firmware/App/Services/Reservoirs.c
DEN-11750 updated the files with software configurations

    • -2
    • +2
    /firmware/App/Controllers/DGInterface.c
    • -17
    • +18
    /firmware/App/Controllers/PresOccl.c
    • -4
    • +10
    /firmware/App/Controllers/Voltages.c
    • -4
    • +12
    /firmware/App/Services/Reservoirs.c
Merge branch 'DEN-11250-hd-syringe-pump-updates' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-11250-hd-syringe-pump-updates

DEN-11750: Updated BP occlusion sensor thresholds.

DEN-11250 Fixed function getHeprinBolusTargetRate declared implicitly

    • -0
    • +1
    /firmware/App/Controllers/SyringePump.h
DEN-11250 Fixed occlusion alarm did not triggered. ForceAtEndOfPrime = 0.6V with 1 second delay, ForceAtEndOfPrime = 2.5V with 3 seconds delay

DEN-11250 Do not trigger syringe pump running while BP is off alarm #83 when syringe pump is in heprin bolus state

    • -3
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-11250 Fixed syringe pump does not retract DEN-11824 and removed redundant code in handleWetSelfTestStartState

DEN-11750 added software configurations and added the DVT arterial pressure sensor

    • -31
    • +37
    /firmware/App/Controllers/Bubble.c
    • -14
    • +17
    /firmware/App/Controllers/DGInterface.c
    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
    • -14
    • +64
    /firmware/App/Controllers/PresOccl.c
    • -1
    • +3
    /firmware/App/Controllers/SyringePump.c
    • -1
    • +1
    /firmware/App/Controllers/Temperatures.c
  1. … 6 more files in changeset.
DEN-11250 Implemented heprin bolus target rate override

    • -3
    • +68
    /firmware/App/Controllers/SyringePump.c
    • -0
    • +2
    /firmware/App/Controllers/SyringePump.h
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Enabled a build switch

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Disabled the temepratures alarm

    • -1
    • +1
    /firmware/App/Controllers/Temperatures.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop