DEN-14834 changed one of the pre-treatment target fills from 600 to 750mL Changed the reservoir's upper limit in treatment from 1950 to 1900mL

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Address comment

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflicts.

DEN-13154: Removed dry self-test timeout alarm. Fixed bug DEN-14884.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Removed obsolete pressure data from broadcast. Tightened no cartridge pressure range for PBA and PBo.

DEN-14834 added code to calculate the heating parameters in pre-treatment and send it to DG in the pre-treatment reservoir management

DEN-14830: reworked dialysate temp timeouts and conditions. Restored previous persistence alarms. added below 33 degC alarm

Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

HD-DEN-14457-2_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Checkpoint. Simplifying AIR_PUMP bug fix.

DEN-14689: synchronized naming from msgdefs.h

DEN-14689: DEN-14634: fixed HD_POST_FINAL_TEST_RESULT being sent to be sent 1 time instead of 40 times over 2s.

DEN-14689: further msg id name changing.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14689-messaging-bugs' into develop

Merge branch 'staging' into DEN-14689-messaging-bugs

DEN-14689: message naming sync

DEN-14689: naming consistency pass.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflicts.

DEN-13154: Update to handling of insufficient Heparin alarm per DEN-14814 and DEN-14287.

DEN-14860 Use higher percentage limit of depletion time alarm for faster dialysate flow. Linear interpolation between fast and slow reservoir fill speeds. Heater ON during reservoir transfer.

DEN-14860 Use higher percentage limit of depletion time alarm for faster dialysate flow. Linear interpolation between fast and slow reservoir fill speeds. Heater ON during reservoir transfer.

DEN-14860 Use higher percentage limit of depletion time alarm for faster dialysate flow. Linear interpolation between fast and slow reservoir fill speeds. Heater ON during reservoir transfer.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14689-messaging-bugs' into develop

    • -110
    • +99
    ./Services/SystemCommMessages.c