DEN-5381: Wait for blood pump finished homing before transition to prime

DEN-5381: Reset pre-treatment state machine and deleted unused code

DEN-5381: Moved DG start and valves homing into pre-treatment mode

DEN-5674: Fixed alarm audio volume control per HDD change.

DEN-5831: Removed eval board and unused includes

DEN-5381: Moved new alarms to available slots

DEN-5381: Increased filling volume to 1700 mL

DEN-5381: Updated hd to dg drain cmd

DEN-5381: Removed reservoir drain and fill from standby mode

DEN-5674: Pass valves check when disable valves build switch is on.

DEN-5674: Added handling for resume after alarm. Removed obsolete build switches.

  1. … 5 more files in changeset.
DEN-5674: Fix alarm state bug.

DEN-5674: Added alarm silence handling.

DEN-5674: Moved alarm lamp self test from early to later in sequence.

DEN-5674: temporary set alarm volume low for testing.

DEN-5674: Fixed alarms state update.

DEN-5674: Prevent resume/rinseback options if top alarm condition not yet cleared.

DEN-5381: Added reservoir management to prime sub-mode

DEN-5674: Completed UF volume setting pre-treatment.

DEN-5674: Removed obsolete test code. Added new UF param msg handler.

  1. … 3 more files in changeset.
DEN-5381: Refactor alarm management module

DEN-5381: Added time out for purge air state

DEN-5674: Fixed valves POST call. Added alarm for FPGA comm timeout.

DEN-5381: Added handler for prime sub-mode states

DEN-5674: Removed alarm dialyzer bypass flag. Added alarm user option for alarms that simply require acknowledgement. Added some alarm action handling for treatment mode.

DEN-5381: Created skeleton for pre-treatment mode

DEN-5674: Added mode specific masking of enabled user alarm recovery actions.

DEN-5674: Fixed drain reservoir cmd to DG to include indication of whether reservoir load cells should be tared.

DEN-5674: Added alarm (empty) action signal handling functions to the various op modes.

  1. … 5 more files in changeset.
DEN-5674: Added alarm sub-rank property and use for determining top priority alarm for UI.