Fixed merge conflict - missing new alarm IDs.

Fixed merge conflict from common.

Fixed merge conflicts from master. Fixed some dev testing issues from merge.

  1. … 12 more files in changeset.
DEN-4641: Fixed PBA conversion formula.

DEN-4641: Pull safety line when cannot stop pumps.

DEN-4641: Added support for Dialin request of calibration data.

DEN-4641: Changed HD pump Dialin set point override commands to simply set commands.

DEN-4641: Removed last change allowing op mode restore from fault mode.

DEN-4641: More fixes to HD pump motor speed measurement. Dialin cmd to clear all alarms will also restore op mode to prior if in fault mode.

DEN-4641: Minor fixes to pump speed measurement and Dialin alarm features per verification.

DEN-4641: Calculate HD pump motor speed more frequently. Added support for Dialin command to clear all alarms.

DEN-4817: Changed alarms' behaviors

DEN-4640: Updated HD pump alarm properties per SRS.

DEN-4640: cosmetic changes.

DEN-4640: clear pump direction error counter if not in control state.

DEN-4640: Added persistence to pump direction and blood pump max rotor speed alarms.

DEN-4640: Updated IT.

DEN-4640: Updated UT. Fixes per UT.

DEN-4640: code review.

DEN-4640: Addressed code review comments.

DEN-4640: Fix disable valves build flag. UT updates.

DEN-4640: Amended response msg to UF pause/resume request from UI.

DEN-4641: Doxygen comment fixes.

DEN-4641: minor fixes.

DEN-4641: Added pressure alarms. Changed occlusion data type to integer and unit from mmHg to unitless.

DEN-4640: Build switch fix.

DEN-4641: Updated inline pressure conversions.

DEN-4641: Fixed PBA conversion to mmHg.

DEN-4640: Addressed code review comments.

DEN-5053: Addressed code review comments.