Merged DEN-14175

DEN-14175 replaced the V3 flag with FPGA hardware check and added the default the fixed valves PWM count in FPGA

0.0F

0.0F

0.0F

0.0F

0.0F

0.0F

DEN-14170 Fix TODO with new fault.

HD-DEN-14170_SW Dev Sprint 82 Darren
HD-DEN-14170_SW Dev Sprint 82 Darren
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Update function header.

Update function header.

Remove blank line

Remove blank line

DEN-14175 addressed code review comments

Done.

Done.

This function should stay commented. I have a TODO task for this.

This function should stay commented. I have a TODO task for this.

Done.

Done.

No these should be TRUE.

No these should be TRUE.

DEN-14175 updated the reservoir switch command to send the trimmer use the previous duty cycle or estimate the duty cycle

Add a space between "return" and "(".

Add a space between "return" and "(".

Update DialInFlow.c, DGInterface.c

DEN-14150: Fixed blending logic for dialysate flow estimation.

DEN-14170 Remove TODO, no longer relevant.

Update DialInFlow.c

DEN-14150: Updated DialInFlow.c

Update DialInFlow.c

DEN-14150: Updated DialInFlow.c

Merge with DEN-14150

DEN-14150: Fix issue with flow estimation blending at 300 mL/min.

DEN-14175: Added f/w support for alarm dialog blood stopped countdown while in rinseback state per DEN-10791.

DEN-14175: Additional fix to syringe pump DAC write issue.

DEN-14175: Stop syringe pump on transition to fault mode only. Continuous stopping of pump was preventing DAC write on force sensor check failure. DEN-14254.