Merge branch 'staging' into DEN-14763-dg_hd_dev-hd_dg_dvt-update-part-10

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14834 updated the code to control the primary heater

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Further fix to DEN-14814 re: do not require syringe removal to re-do preload operation.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14834 changed one of the pre-treatment target fills from 600 to 750mL Changed the reservoir's upper limit in treatment from 1950 to 1900mL

Fixed.

Fixed.

Fixed.

Fixed.

Done.

Done.

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Address comment

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. D...

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm.
And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm.
Did Systems want to separate these? My understanding was not to separate.

Add blank line before return.

Add blank line before return.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflicts.

DEN-13154: Removed dry self-test timeout alarm. Fixed bug DEN-14884.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Removed obsolete pressure data from broadcast. Tightened no cartridge pressure range for PBA and PBo.

DEN-14834 added code to calculate the heating parameters in pre-treatment and send it to DG in the pre-treatment reservoir management