firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Updated signalAllowDGFlushFills(),

signalAllowDGFillRes[12]() to be sure that the signal to continue occurs even when both thr air pump and dry tests are disabled.

Add timeWaitToFillMS to header.

Add timeWaitToFillMS to header.

Aligned memory with sector boundaries

Aligned memory with sector boundaries

Aligned memory with sector boundaries

HD-DEN-14788_DIAL_242 Air Bubble Sensor
HD-DEN-14788_DIAL_242 Air Bubble Sensor
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-14788: DIAL-242: added check for bubble sensor override.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-14906: Added delay before resuming dialyzer re-prime to allow DPi to stop to prevent pressure spike.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
DEN-14888: Fix issue with initializing start/final reservoir volumes.

Merge branch 'DEN-14892-add-bootloader-to-existing-dg-firmware' of ssh://dvm-linux02:7999/hd/hdfirmware into origin/DEN-14892-add-bootloader-to-existing-dg-firmware

    • -180
    • +192
    ./App/Services/SystemCommMessages.c
    • -2
    • +76
    ./App/Services/SystemCommMessages.h
Added project file

Added project file

Added project file

Rearrange stack location

Rearrange stack location

Rearrange stack location

DEN-14906: Handle UF pause and auto-resume for dialyzer re-prime.

I think we need to re-visit this. Could impact dialysate heating negatively.

I think we need to re-visit this. Could impact dialysate heating negatively.