DEN-15183: Fixed DEN-15198 by not starting air trap control until treatment starts (patient connected).

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14325 addressed the code review comments

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

It is not needed. So I removed it.

It is not needed. So I removed it.

Needed?

Needed?

DEN-15183: Addressed code review comments. Fixed DEN-15194.

Fixed.

Fixed.

Fixed.

Fixed.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

DEN-15183: Implemented updates to treatment recirculation workflow to include states for instructing user to disconnect and reconnected patient. Updated scalar for alarm audio current high gain signal. Updated alarm audio POST.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit

DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

Add treatmentTimeBroadcastTimerCtr to header.

Add treatmentTimeBroadcastTimerCtr to header.

DEN-15024 Change syringe pump off error count to max time windowed count.

Add doorOpened to header.

Add doorOpened to header.

DEN-15183: indicate treatment over in treatment times broadcast if alarm preventing alarm resumption.

HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15196_Alarm 98 Rework
HD-DEN-15196_Alarm 98 Rework
DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

DEN-15104 Minor correction

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adding an alarm blocking state to prevent cascade of errors during power fail.

HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming
HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
DEN-15183: Addressed DEN-15191 by zeroing treatment time remaining in treatment times broadcast msg if an alarm has blocked resumption of treatment.