Merge branch 'DEN-15024-fw-srshd-1333-vtchd-283-alarm_id_hd_syringe_pump_not_stopped_error-id-111-does-not' into develop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Switched to CPLD power loss function, rather than alarm.

Add alarmStatus to header

Add alarmStatus to header

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

DEN-15183: Prevent alarm user options while AC power is out.

Add testStartTime, audioTestStartTime, alarmAudioTestToneRequested and to header

Add testStartTime, audioTestStartTime, alarmAudioTestToneRequested and to header

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Moved block to choke point in activateAlarm2Data. Bock all alarms, rather than sub-rank 10 and above.

Extra blank line.

Extra blank line.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15183: Fixed DEN-15198 by not starting air trap control until treatment starts (patient connected).

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14325 addressed the code review comments

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

It is not needed. So I removed it.

It is not needed. So I removed it.

Needed?

Needed?

DEN-15183: Addressed code review comments. Fixed DEN-15194.

Fixed.

Fixed.

Fixed.

Fixed.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

DEN-15183: Implemented updates to treatment recirculation workflow to include states for instructing user to disconnect and reconnected patient. Updated scalar for alarm audio current high gain signal. Updated alarm audio POST.

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit

DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C